Responding to Alarms
42
If this light is
on
And this condition
exists
It means...
Loose Lead
Continuous yellow
light and continuous
alarm
The yellow loose lead light and the alarm may sound
continuously when there is a problem with any of the following:
Lead wires
Electrodes
Electrode belt
Patient cable, or
Connections between the baby’s skin and the electrodes,
the lead wires, the patient cable, and the SmartMonitor 2.
Loose Lead
Continuous yellow
light and no alarm
If you correct the problem, the alarm will stop. However, the
yellow light remains on until you press the
RESET
button.
Power Continuous
green
light, continuous
alarm, with no other
lights lit.
Check the display for error messages. If no error messages
are displayed, the SmartMonitor 2 was turned off improperly
causing a sibling alarm.
To resolve:
Press and hold the blue RESET button.
Press and release the gray POWER button.
Wait 2 seconds then release the RESET button.
All
All lights are blinking
and the alarm will
come on for 3 seconds
and then off for 1
second.
Check the display for error messages. If there is an error
message, enter it on your log sheet.
Turn the monitor off and then back on. If the monitor functions
normally continue to use the monitor.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...