Reducing False Alarms
43
Reducing False Alarms
Proper electrode placement will minimize false alarms.
Make sure the electrodes are placed along the mid-line
of the side, two finger widths below or lined up with the
nipples.
If using the black reusable electrodes with the Velcro
belt, ensure the belt is quite snug. Place the electrodes
far enough apart so that when the belt is wrapped
around the baby, the electrode will be located along the
mid-line of the side, two finger widths below or lined up
with the nipples.
The skin should be clean and dry. If the skin is
unusually dry you may add a few drops of moisture
(water) to the baby’s skin prior to electrode belt
placement.
When using the black reusable electrodes ensure that
the electrode surface is clean.
The optional green (RL) electrode and lead wire may
help reduce EMI interference. Place the green
electrode along the outside of baby’s upper thigh.
Baby’s
right side
Baby’s
left side
The white lead wire
location is illustrated with a
white
electrode.
Refer to the Synergy-E Manual for more information on how to check signal quality using
a modem or using a direct-connect cable.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...