Home Setup
37
Disposable Self Adhesive Electrodes
Follow the steps below if you are using disposable electrodes.
Attach lead wire to the Self Adhesive Electrodes if not pre-attached.
Ensure the infant’s skin is clean and dry.
Place the electrode with the
white
lead wire on the baby’s
right
side, along the mid-line of the side,
two finger widths below or lined up with nipples.
Place the electrode with the
black
lead wire on the baby’s
left
side, along the mid-line of the side, two
finger widths below or lined up with nipples.
An electrode belt is not needed when using disposable electrodes
Use of the third (green - RL) electrode and lead wire is normally not required, but may help
reduce EMI interference. Place the green electrode along the outside of baby’s upper thigh.
Do not use oils, lotion, or powder on the area of skin that the electrodes will be placed.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...