Intro
Read This
First
Transmitter
Installation
Power Supply
Wiring
I/O
Wiring
Batch
Wiring
Startup
Trouble-
shooting
Program-
ming
Main
Menu
Config
Menu
System
Menu
Model
Numbers
Xmitter
Specs
Return
of Goods
Calibration
Menu
Diagnostic
Menu
Datamate 2200 O&M
Page 91
Menu or Display
Description
Configure Flow Vars
Press ENTER to configure Net Oil Functions. This menu will be active only if Datamate 2200
5. Net Oil Flow
was purchased with the Net Oil and Well Testing option. It is important to input first
the Well Data, before attempting to configure the Net Oil functions (please refer to the
Configure Xmit Vars, 5. Well Data
menu found below).
Use the UP or ACROSS key to select if the flow rates and totals are going to be
displayed and calculated in Mass or Volume units.
Net Mass or Volume
[XXXXXXXXXX]
Select the Well number that will be used, this will provide the density of dry crude oil
and saline water used in the calculations.
Use Data from Well
[XXX]
To configure the Water Cut, Net Oil and Net Water measurement, select the
corresponding option using the UP or ACROSS key and then pressing ENTER.
Net Oil Parameter
[XXXXXXXXXXXXXXXXXX]
Use the UP or ACROSS key to select unit.
Unit
[XXXX
]
Use the UP or ACROSS key to set the fallback value in the event of a fault
condition
Fallback
[XXXXXXXXXXXX]
Use the UP or ACROSS key to set the visual Low Alarm Setpoint.
Low Alarm SetPnt
[XXXXXXXXXXXX]
Use the UP or ACROSS key to set the visual High Alarm Setpoint.
Hi Alarm SetPnt
[XXXXXXXXXXXX]
Only in the cases of Net Oil and Net Water parameter will be displayed the
following prompt. Use the UP or ACROSS key to set the Low Flow Cutoff
point. Below this value the meter will not register the flow rate.
LF Cutoff
[XXXXXXXXXXXX]
Only in the case Net Oil is selected as a parameter, it is necessary to provide
the following variable, the allowable range for this field is between 0.5 and
1.5. The factory default is 1.0 (please refer to the
chapter 7.11 Net Oil and
Well Testing
for an explanation on the meaning of this field).
Shrinkage Factor
[XXXXXXXXXXXX]