SECTION 12: REPLACING A DM2100 WITH A DM2200
Due to parameter changes from the DM2100 to the DM2200, the following information
should be entered into the DM2200 upon changing out.
• Refer to Pg. 58 of the DM2200 O&M Manual for correct path to Electronic Zero, RTD
Offset, Density Calibration Coefficients, and Scalar.
• Refer to Pg. 79 for Low-Flow Cutoff (Under Xmit Variables, Flow Variables, Mass Flow)
• The average values are to be entered into the DM2200 prior to test run.
Note: The relays supplied with the DM2200 are for AC systems only and will not operate
on DC voltage. Relays for DC systems must be supplied by the user.
Meter Size
Scalar
Average
Density Calibration
Coefficients
Low-Flow
Cutoff
(lb./min.)
Average
RTD Offset
Average
Elect. Zero
C1
C2
1/8"
6.3528E+03 1.42E+05
1.53E+01
0.1
-0.0257
4.42E-05
1/4"
3.9242E+04 4.49E+04
6.05E+00
0.4
-0.0257
4.42E-05
1/2"
8.8785E+04 3.47E+04
3.84E+00
1.5
-0.0257
4.42E-05
1"
2.6172E+05 2.83E+04
3.45E+00
5.5
-0.0257
4.42E-05
2"
7.2392E+05 2.62E+04
2.59E+00
16
-0.0257
4.42E-05
3"
1.7596E+06 6.18E+03
1.88E+00
45
-0.0257
4.42E-05
4"
1.5633E+06 8.70E+03
2.53E+00
75
-0.0257
4.42E-05