Page 68
9.5.5 Analog Output Calibration
NOTE: None of the analog outputs will function until they are programmed.
Using the CALIBRATION MENU TREE enter ANALOG OUTPUT MENU
The procedure for the programming of the analog outputs is as follows:
Menu or Display
Description
Calibration
From this menu the calibration factors for the analog (4-20 mA) outputs
6. Analog Outputs
can be viewed and modified.
Choices from this menu: Analog 1, Analog 2, Analog 4, & Exit.
Analog Outputs
Choices from this menu are Analog 1 Params, Calibrate Analog 1,
1. Analog 1
and Exit.
Calibrate Analog 1
From the Analog 1 Params menus the user can view and manually
1. Analog 1 Params
change (with level 2 access) analog 1’s calibration factors.
Pressing ENTER displays the Gain factor
Gain
[_X.XXXXeSXX]
Offset
[_X.XXXXeSXX]
Press CANCEL to Exit when complete.
Calibrate Analog 1.
From this menu the Datamate 2200 will re-determine the calibration
factors with Calibrate 1 Analog the assistance of the user. The user is
required to measure and enter the output current when directed.
Pressing Enter will bring up:
Cal Current 1 = 4.000 mA
Enter current measured at Analog Output 1
[_X.XXX] mA
Cal Current 2 = 19.000 mA
Enter current measured at Analog Output 1
[_XX.XXX] mA
Gain
[_X.XXXXeSXX]
Offset
[_X.XXXXeSXX
Analog Outputs
See Analog Outputs
2. Analog 2
1. Analog 1 above.
Calibrate Analog 2
See Calibrate Analog 1
1. Analog 2 Params
1. Analog 1 Params above.
Calibrate Analog 2
See Calibrate Analog 1
2. Calibrate Analog 1 1. Calibrate Analog 1 above
Analog Outputs
See Analog Outputs
3. Analog 4
1. Analog 1 above.
Calibrate Analog 4
See Calibrate Analog 1
1. Analog 4 Params
1. Analog 1 Params above.
Calibrate Analog 4
See Calibrate Analog 1
2. Calibrate Analog 1 1. Calibrate Analog 1 above
Calibration
7. Exit