Instrument functions
R&S
®
NGU
73
User Manual 1179.2531.02 ─ 03
15. Depending on your requirement, select "Start triggered" or "Step triggered" menu
item and set "On".
The R&S
NGU starts or steps through the arbitrary signal when a trigger is detec-
ted.
6.10
Advanced features
Arbitrary function
If Arbitrary function is enabled, the channel voltage, current setting and safety limit set-
tings are disabled.
See
Chapter 6.1, "Setting the channels voltage and current"
The Arbitrary and Ramp are two functions which can be used to control the waveform
of voltage and current output.
6.10.1
Arbitrary
The Arbitrary function allows you to generate freely programmable waveforms which
can be reproduced within the
1. Press [Settings] key.
The R&S
NGU displays the device/channel menu window.
2. Select the channel tab to configure Arbitrary function.
The R&S
NGU displays the channel menu.
3. Select "Arbitrary" from the menu.
The R&S
NGU displays the "Arbitrary" dialog.
Figure 6-39: Arbitrary dialog
Advanced features