Remote control commands
R&S
®
NGU
140
User Manual 1179.2531.02 ─ 03
SENSe:CURRent:RANGe:AUTO:UP:THReshold
<arg0>
Sets or queries the threshold duration before the instrument switches the measure-
ment range when an overrange condition is detected.
Parameters:
<arg0>
Range:
2
µ
s to 0.13 s
*RST:
20
µ
s
Example:
SENS:CURR:RANG:AUTO:UP:THR 0.001
The instrument does not switch the measurement range until the
overrange condition exceeds the 0.001 seconds.
SENSe:CURRent:RANGe:AUTO
<arg0>[, <Channel list>]
SENSe:CURRent:RANGe:AUTO?
[<Channel list>]
Sets or queries auto range for current measurement accuracy.
Parameters:
<arg0>
1
Enables auto range for current.
0
Disables auto range for current.
Parameters for setting and query:
<Channel list>
<list>
Example:
SENS:CURR:RANG:AUTO 1
Enables auto range for current.
Example:
SENS:CURR:RANG:AUTO? (@1)
Queries auto range state for current measurement accuracy at
channel 1.
SENSe:CURRent:RANGe[:UPPer]
<arg0>[, <Channel list>]
SENSe:CURRent:RANGe[:UPPer]?
[<Channel list>]
Sets or queries the current range for measurement. There is a selection of 10 A, 3 A,
100 mA, 10 mA, 1 mA and 10
µ
A range.
Parameters:
<arg0>
Defines the current range for measurement.
Default unit: A
Parameters for setting and query:
<Channel list>
<list>
Example:
SENS:CURR:RANG 10
Sets the instrument to the 10 A measurement accuracy.
Refers to datasheet for the measurement accuracy in the 10 A
range.
Configuration commands