[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
............................................. 126
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
....................................................... 126
SOURce:CURRent:NEGative[:LEVel][:IMMediate][:AMPLitude]
.......................................... 127
...........................................................................................128
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
<stepsize>
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]?
[<Optional default
step query>]
Sets or queries the incremental step size for the
Setting parameters:
<stepsize>
<numeric value> | DEF | DEFault
<numeric value>
Step value in A.
DEF | DEFault
Default value of stepsize.
Range:
0.0001 to 2.000
Increment: 0.0001
*RST:
0.010
Default unit: A
Parameters for setting and query:
<Optional default step
query>
DEFault
Queries the default voltage step size.
Example:
CURR:STEP 0.005
CURR:STEP DEF
VOLT:STEP? DEF
-> 0.1000E+00
Returns the default stepsize for current.
See
Example "Configuring the current output"
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
<current>
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
[<Channel list>]
Sets or queries the current value of the selected channel.
Parameters:
<current>
<numeric value> | MIN | MINimum | MAX | MAXimum | UP |
DOWN | <list>
<numeric value>
Numeric value in the range of 0.000 to 6.010.
MIN | MINimum
Minimum current at 0.010 A.
MAX | MAXimum
Depending on the set voltage level, the maximum set current is
6.010 A.
For voltage range from 0 V to 6 V, maximum set current is 6.01
A.
For voltage > 6 V, maximum set current is 3.01 A.