12.2 Preset Alarm Limits & Calibration
The instrument is factory calibrated with standard calibration gas, and
is programmed with default alarm limits.
Cal Gas
Cal
Span
unit
Low High TWA STEL
Isobutylene 100 ppm 50 100 10 25
Benzene 5
ppm
2 5 0.5
2.5
Butadiene 10 ppm
5 10 2 5
12.3 Testing The Alarm
You can test the alarm whenever the main (Reading) display is
shown. Press [Y/+], and the audible and visible alarms are tested.
13 Integrated Sampling Pump
The instrument includes an integrated sampling pump. This
diaphragm-type pump that provides a 450 to 550 cc per minute flow
rate. Connecting a Teflon or metal tubing with 1/8" inside diameter to
the gas inlet port of the instrument, this pump can pull in air samples
from 200' (61 m) away horizontally, or 90' (27.5 m) vertically, at
about 3' (0.9 m) per second flow speed.
Note:
In Search Mode, the pump turns on when a sample
measurement is started, and turns off when the sample is manually
stopped.
If liquid or other objects are pulled into the inlet port filter, the
instrument detects the obstruction and immediately shuts down the
pump. The alarm is activated and a flashing pump icon is displayed.
You should acknowledge the pump shutoff condition by clearing the
obstruction and pressing the [Y
/+
] key while in the main reading
display to restart the pump.
43
Summary of Contents for UltraRAE 3000
Page 31: ...UltraRAE 3000 User s Guide 29...
Page 41: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 39...
Page 59: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 57...
Page 79: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 77...
Page 95: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 93...
Page 97: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 95...
Page 128: ......