UltraRAE 3000 User’s Guide
10.5 Inserting The Separation Tube
1.
Unscrew the front of the sampling probe from the base.
2.
Slip the tube into the rubber holder in the front portion. Make
sure the arrow on the side of the tube points toward the
instrument.
3.
Insert the other end of the tube into the middle of the base while
turning the front portion to tighten it onto the base’s threads.
IMPORTANT!
Do not overtighten any portion of the sampling assembly.
Note:
When the UltraRAE 3000 is used for VOC monitoring, no tube
is inserted.
IMPORTANT!
To ensure that there are no leaks, periodically test the seals:
With the UltraRAE 3000 running, place your finger over the end of
the inlet probe. The alarm should sound and the pump-stall icon
should flash on the display. This indicates that all seals are good. Stop
the alarm by pressing [Y/+]. If the pump does not alarm or show the
stalled-pump icon, then check that all inlet parts are tight and inspect
the O-ring for damage (replace it if necessary).
34
Summary of Contents for UltraRAE 3000