UltraRAE 3000 User’s Guide
25.4 Diagnostic Mode
IMPORTANT!
Diagnostic Mode is designed for servicing and
manufacturing, and therefore is not intended for everyday use, even
by advanced users. It provides raw data from sensors and about
settings, but only allows adjustment of pump stall parameters, which
should only be changed by qualified personnel.
Note:
If the instrument is turned on in Diagnostic Mode and you
switch to User Mode, datalog data remains in raw count form. To
change to standard readings, you must restart the instrument.
25.4.1 Entering Diagnostic Mode
Note:
To enter Diagnostic Mode, you must begin with the instrument
turned off.
Press and hold [Y/+] and [MODE] until the instrument starts.
The instrument goes through a brief startup, and then displays raw
data for the PID sensor. These numbers are raw sensor readings
without calibration. The instrument is now in Diagnostic Mode.
Note:
In Diagnostic Mode, the pump and lamp are normally on.
You can enter Programming Mode and calibrate the instrument as
usual by pressing both [MODE] and [N/-] for three seconds.
You can enter Monitoring Mode by pressing [MODE] and [Y/+]
together for three seconds.
Once the instrument is started up in Diagnostic Mode, you can switch
between Diagnostic Mode and Monitoring Mode by pressing and
holding [MODE] and [Y/+] simultaneously for two seconds.
In Diagnostic mode, you can step through parameter screens by
pressing [MODE].
97
Summary of Contents for UltraRAE 3000
Page 31: ...UltraRAE 3000 User s Guide 29...
Page 41: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 39...
Page 59: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 57...
Page 79: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 77...
Page 95: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 93...
Page 97: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 95...
Page 128: ......