UltraRAE 3000 User’s Guide
27 Maintenance
The major maintenance items of the instrument are:
•
Battery pack
•
Sensor module
•
PID lamp
•
Sampling pump
•
Inlet connectors and filters
In addition, you can test the humidity sensor (this should be done
annually, in order to ensure the most accurate operation).
Note: Maintenance should be performed by qualified personnel
only
.
NOTE: The printed circuit board of the instrument is connected
to the battery pack even if the power is turned off
.
Therefore, it is
very important to disconnect the battery pack before servicing or
replacing any components inside the instrument
.
Severe damage
to the printed circuit board or battery may occur if the battery
pack is not disconnected before servicing the unit.
27.1 Battery Charging & Replacement
When the display shows a flashing empty battery icon, the battery
requires recharging. It is recommended to recharge the instrument
upon returning from fieldwork. A fully charged battery runs a
instrument for 16 hours continuously. The charging time is less than 8
hours for a fully discharged battery. The battery may be replaced in
the field (in areas known to be non-hazardous), if required.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous
.
Remove and
replace battery only in areas known to be non-hazardous.
104
Summary of Contents for UltraRAE 3000
Page 31: ...UltraRAE 3000 User s Guide 29...
Page 41: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 39...
Page 59: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 57...
Page 79: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 77...
Page 95: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 93...
Page 97: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 95...
Page 128: ......