1. Potentiometric measuring principles
ABL800 FLEX Reference Manual
Electrolyte electrodes,
Continued
Stability criteria
The following stability criterion must be met to obtain a stable electrode response
during calibration:
upd.last)
X(Cal,
K
)
upd.i
X(Cal,
upd.last)
X(Cal,
c
c
c
×
≤
−
This criterion is valid for calibrations using Cal 1 and Cal 2 solutions where:
c
X(Cal,upd.last) =
Concentration of the electrolyte ion from the last updating
when measuring on calibration solution. (The last
updating is number 30).
c
X(Cal,upd.i) =
Concentration of the electrolyte ion for a given updating
when measuring on calibration solution. (The relationship
must be fulfilled for at least one of the updating numbers
18 or 19).
K =
Constant for the stability criterion.
Electrolyte Ion
Cal1 solution
Cal2 solution
K
+
0.01 0.01
Na
+
0.01 0.02
Ca
2+
0.02 0.02
Cl
−
0.022 0.022
The following stability criterion must be met to obtain a stable electrode response
during measurement:
(
)
)
(
)
(
)
upd.last
X(sample,
K
)
upd.i
X(sample,
upd.last)
X(sample,
Rinse
cX
Rinse
cX
c
c
c
+
−
×
≤
−
where:
c
X(sample,upd.last) =
Concentration of the electrolyte ion from the median
of the last 5 updatings (for Ca
2+
: 3 last updatings)
when measuring on a sample. The last updating
number is 30 (or 10 for some micromodes).
c
X(sample,upd.i) =
Concentration of the electrolyte ion for a given
updating when measuring on a sample. (The
relationship must be fulfilled for at least one of the
updating numbers shown below).
K
+
Na
+
Ca
2+
Cl
−
22
22
26
22
23
23
27
23
In some micromodes, substract 20 from number
above.
Continued on next page
1-32