ABL800 FLEX Reference Manual
2. Amperometric measuring principles
Metabolite electrodes,
Continued
Stability criteria
(continued)
If all the criteria below are fulfilled, then the result of the measurement will be
marked with an interference error.
1
upd.9)
I(sample,
-
upd.16
I(sample,
upd.23)
I(sample,
-
upd.30)
I(sample,
≥
I(sample,upd.16) > I(sample,upd.12)
I(sample,upd.12) > I(sample,upd.9)
c
X(sample,corr)
>
1.5 mmol/L
where:
I(sample,upd.30)
I(sample,upd.23)
I(sample,upd.16)
I(sample,upd.12)
I(sample,upd.9)
= Electrode current at the 30
th
/23
rd
/16
th
/12
th
/9
th
updating during measurement on sample,
respectively.
c
X(sample,corr)
= Corrected concentration of glucose or lactate in
the sample.
2-21