5. Performance characteristics
ABL800 FLEX Reference Manual
Performance test results – cGlu
Primary
reference
method
Spectrophotometry, using the hexokinase (HK) method recommended by NCCLS
[
5
]
, measured on serum.
Bias
REF
The FLEXMODE on the ABL805/35 analyzers was tested:
cGlu(mmol/L)
Bias
REF
N
0.08 0.03
30
2.09 0.06
30
5.08 0.12
30
14.73 -0.02
30
N = number of measurements on several analysers used for the test.
Bias
ABL
– blood
samples
This bias is presented by the following chart:
Bias (%)
ABL735 (mmol/L)
Continued on next page
5-24