ABL800 FLEX Reference Manual
5. Performance characteristics
Interference tests
pH/blood gas
The following interference results are found for the pH and blood gas electrodes:
Substance Test
Conc. Interference
on
pO
2
Electrode
Halothane
3 %
5 % increased sensitivity
Intralipid (20 % solution) in a concentration greater than 4 % (the final Intralipid
level being 0.8 %) will give interference on pH measurements.
Electrolytes
The following interference results are found for the electrolyte electrodes:
Interference
on…
Substance Test
Conc.
cK
+
(4 mmol/L
level)
cNa
+
(150 mmol/L
level)
cCa
2+
(1.25 mmol/L
level)
cCl
−
(110 mmol/L
level)
Li
+
4 mmol/L
0
0
0
K
+
12 mmol/L
−
1
−
0.01
Na
+
100 - 180
mmol/L
0.1 to
−
0.1
NH
4
+
1 mmol/L
0
0
Ca
2+
5 mmol/L
0
Mg
2+
5 mmol/L
0
0
0.05
Br
−
10 mmol/L
41
F
−
1 mmol/L
0
I
−
3.0 mmol/L
30-90
ClO
4
–
1.5 mmol/L
8-30
HCO
3
–
25-50 mmol/L
0.1 mmol/L Cl
−
per mmol/L
HCO
3
–
Lactate 10
mmol/L
0
Acetyl-
salicylic acid
3.0 mmol/L
2
Continued on next page
5-47