5. Performance characteristics
ABL800 FLEX Reference Manual
Performance test results - oximetry,
Continued
FHbF – adult
blood
(continued)
Bias (micromodes):
FHbF
ABL835 ABL830
ctHb
(g/dL)
FHbF
(%)
S95 C95 S85 C55 C35 FM* C55 C35
1
0
0 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3
1
5
0 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5
2
0
0 5.6 5.6 5.6 5.6 5.6 5.6 5.6 5.6
* FM = FLEXMODE (no message)
Imprecision (micromodes):
ctHb (g/dL)
FHbF (%)
sO
2
(%)
S
0
S
D
S
ABL
S
X
10
4 4 5 8
15
2 3 5
7
20
0 100
2 2
10 11
NOTES: a, b.
FHbF – fetal
blood
Bias (macromodes):
FHbF
ABL835
ABL830
FHbF (%)
ctHb (g/dL)
S195
FM*
80 10
5.9 5.9
5.9
80 15
3.3 3.3
3.3
80 20 2.6 2.6
2.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%)
ctHb (g/dL)
sO
2
(%)
S
0
S
D
S
ABL
S
X
80 10 100
4
5
5
9
80 15
100
3
3
6
8
80 20
100
2
3
6
7
NOTES: a, b.
Continued on next page
5-38
Summary of Contents for ABL800 FLEX
Page 198: ...Traceability certificates...
Page 199: ...ABL800 FLEX Reference Manual 7 Solutions and gas mixtures 7 9...
Page 200: ......
Page 201: ...ABL800 FLEX Reference Manual 7 Solutions and gas mixtures 7 11...
Page 202: ......