ABL800 FLEX Reference Manual
5. Performance characteristics
Performance test results – pO
2
Primary
reference
method
Tonometry
[
3
]
.
The gases used for tonometry are traceable to NIST certified Standard Reference
Materials.
Bias
REF
The FLEXMODE on the ABL805/30/35 analyzers was tested:
pO
2
(mmHg)
Bias
REF
N
15 0.47
60
50 -0.24
60
150 -0.45
60
250 -2.17
60
530 1.01
60
N = number of measurements on several analysers used for the test.
Bias
ABL
– blood
samples
This bias is presented by the following chart:
Bias (%)
ABL735 (mmHg)
Continued on next page
5-13