2. Amperometric measuring principles
ABL800 FLEX Reference Manual
Metabolite electrodes,
Continued
Corrections
(continued)
Corrections for cGlu :
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 0.94
0.1
A
S95
1.00
0.0
A,
B
C95
1.06
0.0
A,
B
C35
1.16
0.0
A,
B
*FM (no message)
0.94
0.1
A
*FM (message 874)
1.06
0.0
A, B
*FM (message 873)
1.06
0.0
A, B
05 S165 0.94
0.1
A
S95
1.00
0.0
A,
B
C95
1.06
0.0
A,
B
C35
1.16
0.0
A,
B
*FM (no message)
0.94
0.1
A
*FM (message 874)
1.06
0.0
A, B
*FM (message 873)
1.06
0.0
A, B
*FM = FLEXMODE.
Corrections for cLac:
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 0.97
-0.04
A
S95
1.03
0.03
A,
B
C95
1.03
0.18
A,
B
C35
1.13
0.05
A,
B
*FM (no message)
0.97
-0.04
A
*FM (message 874)
1.03
0.18
A, B
*FM (message 873)
1.03
0.18
A, B
*FM = FLEXMODE.