4. User-defined corrections
ABL800 FLEX Reference Manual
Correction factors for oximetry parameters and bilirubin
Allowed
corrections
The following corrections can be user-defined for the oximetry parameters and
bilirubin:
Parameter
Allowed User-defined Corrections
Slope
Offset
ctHb
Yes No
sO
2
Yes Yes
FCOHb
No Yes
FMetHb
No Yes
FO Hb
2
No No
FHHb
No No
FHbF
Yes Yes
ctBil
Yes Yes
NOTE: In order to define the corrections accurately, the measurements of the
oximetry parameters and bilirubin on the ABL800 FLEX analyzers should be made
without any entered corrections. To avoid truncation errors from an enabled “Out
of range suppression” function it is important to disable the function.
The following recommendations apply to
c
tHb:
ctHb
Item Description
Units
g/dL; g/L; mmol/L
Sample
Set
c
tHb of a SAT100 sample to
≈
15 g/dL (9.3 mmol/L) and
pH
≈
7.4
c
tHb, maximum
point
Uncorrected or corrected:
≈
15 g/dL or 9.3 mmol/L
Slope
0.950 - 1.050
The following recommendations apply to
s
O
2
:
sO
2
Item Description
Units Fraction
Sample
Set
c
tHb of gas equilibrated SAT0 and SAT100 samples to
≈
15 g/dL (9.3 mmol/L) and pH
≈
7.4
Slope
0.900 - 1.100
Offset
±
0.050
Continued on next page
4-4