ABL800 FLEX Reference Manual
2. Amperometric measuring principles
Metabolite electrodes,
Continued
Corrections
(continued)
ABL8XX
FLEX
Mode A
0
A
1
Equation
05 S165 0.97
-0.04
A
S95
1.03
0.03
A,
B
C95
1.03
0.18
A,
B
C35
1.13
0.05
A,
B
*FM (no message)
0.97
-0.04
A
*FM (message 874)
1.03
0.18
A, B
*FM (message 873)
1.03
0.18
A, B
*FLEXMODE = adaptive measuring mode.
Stability criteria
The following stability criteria must be met to obtain a stable electrode response
during calibration:
I(Cal 1,upd.30)
−
I(Cal 1, upd.21)
−
9
×
I
slope
≤
0
S
d,zero
< S
d,max
50
upd.21)
1,
I(Cal
upd.11)
1,
I(Cal
upd.11)
1,
I(Cal
upd.1)
1,
I(Cal
log
5
.
9
≤
−
−
−
=
τ
All of the three criteria must be fulfilled for a calibration using Cal 1 solution
where:
I(Cal 1,upd.30)
I(Cal 1,upd.21)
I(Cal 1,upd.11)
I(Cal 1,upd.1)
= Electrode current at the 30
th
/21
st
/11
th
/1
st
updating during
measurement on Cal 1 solution, respectively.
S
d,zero
= Spreading of the zero point current updatings around the
regression line.
S
d,max
= If Sens
>
400 pA/mM, then
= 0.025
×
Sens,
otherwise
= 10.0.
max
d,
S
max
d,
S
Continued on next page
2-19