5. Performance characteristics
ABL800 FLEX Reference Manual
Performance test results - oximetry,
Continued
sO
2
-
macromodes
Bias:
sO
2
(%)
ABL835/25/15
ABL830/20/10
ctHb (g/dL)
sO
2
(%)
S195
FM*
S85
15 0 0.00 0.05
−
0.02
7 100 0.01 0.22
N/A
15 100 0.01
−
0.08
0.00
25 100 0.00
−
0.29
N/A
* FM = FLEXMODE (no message) corresponding to C195 mode on the
ABL735/25/15.
Imprecision:
ctHb (g/dL)
sO
2
(%)
S
0
S
D
S
ABL
S
X
15 0 0.05
0.05
0.25
0.30
7 100
0.10
0.10
0.25
0.30
15 100 0.05
0.10
0.25
0.30
25 100 0.05
0.10
0.30
0.35
sO
2
-
micromodes
Bias:
sO
2
ABL835/25/15
ABL830/20/10
ctHb
(g/dL)
sO
2
(%)
S95 C95 S85
C55
C35
FM
* C55 C35
15 0
−
0.04
−
0.02
-0.02 -0.03 -0.03 N/A
−
0.03
−
0.03
7 100
−
0.10
−
0.19
N/A -0.22 -0.10 N/A
−
0.22
−
0.10
15 100
−
0.10
−
0.16
0.00 -0.16 -0.10 -0.05
−
0.16
−
0.10
25 100
−
0.10
−
0.17
N/A -0.14 -0.09 N/A
−
0.14
−
0.09
* FM = FLEXMODE (no message) corresponding to C195 mode on the
ABL735/25/15.
Imprecision:
ctHb (g/dL)
sO
2
(%)
S
0
S
D
S
ABL
S
X
15 0
0.05
0.05
0.25
0.30
7 100
0.10
0.10
0.25
0.30
15 100
0.05
0.10
0.25
0.30
25 100
0.05
0.10
0.30
0.35
Continued on next page
5-32