1. Potentiometric measuring principles
ABL800 FLEX Reference Manual
Electrolyte electrodes,
Continued
Corrections for cNa
+
(cont):
Corrections
(continued)
ABL8XX
FLEX
Mode A
0
A
1
Equation
05 S165
0.995
-3.00
A
S95
1.01
1.80
A,
B
C95
1.03
-1.09
A,
B
*FM (no message)
0.995
-3.00
A
*FM (message 874)
1.030
-1.00
A, B
*FM = FLEXMODE.
Corrections for cK
+
:
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 0.985
-0.065
A
S95
1.05
-0.13
A,
B
C95
1.11
-0.37
A,
B
*FM (no message)
0.985
-0.065
A
*FM (message 874)
1.11
-0.37
A, B
05 S165
0.985
-0.065
A
S95
1.05
-0.13
A,
B
C95
1.11
-0.37
A,
B
*FM (no message)
0.985
-0.065
A
*FM (message 874)
1.11
-0.37
A, B
*FM = FLEXMODE.
Continued on next page