MAINTENANCE
/ WEEKLY MAINTENANCE
142
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
12.6.8
CLEANING OF THE MAGAZINE STATION
The magazine station door is controlled by software and cannot be opened when the
MosaiQ 125
is operating, but it can be opened when the
MosaiQ 125
is in
Idle
mode. For
safety reasons DO NOT clean the magazine station door when the
MosaiQ 125
is connected
to the mains power supply.
WARNING
Disconnect the
Mosaiq 125
power cord from the mains power
supply before carrying out this cleaning/disinfection operation.
To clean the magazine station:
1 - Be sure that the
MosaiQ 125
is not connected to the mains power supply.
2 - Open the magazine station door.
3 - Remove any magazine from the magazine station.
4 - Wipe all surfaces of the magazine carousel (all four positions) with a lint free wipe
and 70% alcohol.
5 - Close the magazine station and wipe the external surface of the magazine station
door with a lint free wipe and 70% alcohol.
More information about the magazine station, please refer to
chapter 5.4
of this manual.
To remove magazines, form the magazine station, please refer to
chapter 7.3.1
of this
manual.
12.6.9
CLEANING OF THE COMPUTER WORKSTATION
The computer is located in the water drawer. DO NOT perform the cleaning of the computer
workstation when the
MosaiQ 125
and the computer workstation are connected to the
mains power supply.
WARNING
Disconnect the
Mosaiq 125
power cord from the mains power
supply before carrying out this cleaning/disinfection operation.
To clean the computer workstation:
1 - Be sure that the
MosaiQ 125
and the computer workstation are not connected to the
mains power supply.
2 - Open the water drawer.
3 - Wipe the surface of the computer workstation and the
ON/OFF
button with a lint free
wipe and 70% alcohol.
4 -Close the water drawer.
More information about the computer workstation, please refer to
chapter 5.7.2
of this
manual.
12.6.10
CLEANING OF THE SAMPLE STATION (EXTERNAL)
External part
The external part of the sample station is equipped with a laser barcode scanner and
requires attention when cleaning of this area.
Handling and Cleaning the optical Surfaces:
NOTICE
Improper handling and cleaning of optical surfaces (e.g. scanners,
lenses, sensors) could generally degrade the quality of images,
data, etc.
Do not touch any optical surfaces with anything other than soft lint-free wipes that
can be used for cleaning.
Use a commercially available lens cleaner to clean optical surfaces, never use any
aggressive detergents or solutions (e.g. acetone).
To clean the external part of the sample station:
1 - Wipe the surface with a lint free wipe and 70% alcohol.
12.6.11
CLEANING OF THE TOUCH SCREEN
The touch screen is a sensitive part and it requires attention regarding the cleaning
product to be used.
Damage of touch screen while cleaning
NOTICE
Improper cleaning could damage the touch screen surface.
Use a soft and lint-free wipe with neutral detergent or with ethanol to clean the
touch screen.
Do not use any chemical solvent, acidic or alkali solution.
12.6.12
CLEANING OF THE SURROUNDING FLOOR AREA
Please follow the laboratory internal safety and cleaning instructions.
1 - Any spills must be cleaned up immediately.
2 - Spills containing or potentially containing chemical and/or biological components
must be cleaned up immediately and the area sanitized by spraying 70% alcohol.
3 - All potentially contaminated materials or materials used for cleaning must be
disposed of in the biological waste container.