SAMPLE LOADING AND TESTING
/ SDS TEST RESULTS
104
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
(retesting a sample)
Repeating testing
Repeating testing is required for Syphilis initial reactive result only. For detailed information, please
refer to the MosaiQ SDS Magazine IFU.
There are two different cases for retesting:
භ
The sample is still loaded in the
MosaiQ 125
.
භ
The sample has been removed and is no longer in the
MosaiQ 125
.
The sample is already / still loaded in the MosaiQ 125.
1 - Go to back to the sample loading screen, the sample will automatically be assigned for testing.
2 - Select the appropriate test and tap on the
Add to Process
to run the retesting (as described in
chapter 8.6
)
The sample has been removed and is no longer in the MosaiQ 125.
1 - Load the Sample into the
MosaiQ 125
, the sample will automatically be assigned for testing (as
described in
chapter 8.6
).
Confirmed result after retesting
On the
Report
sub menu of the
Finished
screen you can see details of the tests that require retesting.
1 - Tap on the lines to view more details
(result confirmation after retesting)
This screen shows the
Initial Run
results details and the
1st Confirmation
results details.