MAINTENANCE
/ GUI MAINTENANCE TASKS OVERVIEW
145
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
12.8
GUI MAINTENANCE TASKS OVERVIEW
This section describes the maintenance tasks provided by the GUI.
Procedure
1 - Tap on the
Maintenance
icon in the navigation bar.
2 - Tap on a task to show details.
3 - Tap on the
Execute Task
button to execute the maintenance task.
4 - Follow the instructions.
Tasks
Task
Description
Frequency
Initialize System
Initializes the
MosaiQ 125
.
Daily required after
MosaiQ 125
start
Initial Washer Prime
Primes the wash module 1 and wash
module 2 and generates fresh diluted
wash buffers.
Solutions used:
•
MosaiQ Wash Buffer 1
•
MosaiQ Wash Buffer 2
After
MosaiQ 125
initialization
Clean Pipettor Probes
Cleaning of the pipettor needles
Solution used:
• Sodium Hydroxide
(0.5 M NaOH)
Required once a week
Clean Washer Prime
troughs
Incubation ring priming wells cleaning
(these wells are not visible to users)
Solutions used:
•
MosaiQ Wash Buffer 1
• System
liquid
As needed, follow the
indications on the
MosaiQ 125
Prime Wash Buffer 1
Primes the
MosaiQ Wash Buffer 1
pumps
in the reconstitution drawer. Only if a
reagent bottle is installed.
As needed, follow the
instructions on the
MosaiQ 125
Task
Description
Frequency
Prime Wash Buffer 2
Primes
MosaiQ Wash Buffer 2
pumps
in the reconstitution drawer. Only if a
reagent bottle is installed.
As needed, follow the
indications on the
MosaiQ 125
Shift start
Starts the archiving of results, quality
control results and maintenance tasks.
As needed, follow the
indications on the
MosaiQ 125
Prime Detection Reagent 1
Primes the
MosaiQ Detection Reagent
1
dispensers on the reagent priming
trough. Only if a reagent bottle is
installed.
As needed, follow the
indications on the
MosaiQ 125
Prime Detection Reagent 2
Primes the
MosaiQ Detection Reagent 2
dispensers on the reagent carriage. Only
if a reagent bottle is installed.
As needed, follow the
indications on the
MosaiQ 125
Prime Sample Diluent 1
Primes the
MosaiQ Sample Diluent 1
pumps in the sample diluents reservoir.
Only if a reagent bottle is installed.
As needed, follow the
indications on the
MosaiQ 125
Prime Sample Diluent 2
Primes the
MosaiQ Sample Diluent 2
pumps in the sample diluents reservoir.
Only if a reagent bottle is installed.
As needed, follow the
indications on the
MosaiQ 125
Prime Dispensers
Primes all dispensers (
MosaiQ Detection
Reagent 1
,
MosaiQ Detection Reagent 2
,
MosaiQ Enhancement Reagent 1, MosaiQ
Enhancement Reagent 2
)
on the reagent
carriage.
Only if a reagent bottle is installed.
As needed, follow the
indications on the
MosaiQ 125
Prime Enhancement Re-
agent 1
Primes
MosaiQ Enhancement Reagent 1
dispensers on the reagent carriage. Only
if a reagent bottle is installed.
As needed, follow the
indications on the
MosaiQ 125