OPERATING INSTRUCTIONS
/ HOW TO REMOVE AND LOAD EACH RESOURCE
71
E
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
7.3.4
LOADING DETECTION AND ENHANCEMENT REAGENT
BOTTLES
This section provides step-by-step instructions on how to load reagent bottles (MosaiQ
detection reagents and MosaiQ enhancement reagents) into the
MosaiQ 125
. The system
software controls reagent door locking and unlocking operations.
-
1
Tap on the
Resources
icon in the navigation bar.
-
2
Tap on the
Reagents
area to access the
Reagents
screen.
2
1
(reagents resources area)
-
3
Wait until the
MosaiQ 125
status is
Idle
.
3
(reagents resources detais screen)
NOTICE
To know the remaining time before
MosaiQ 125
goes to
Idle
, go to
the
Results
menu and tap on the
Processing
button.