OPERATING INSTRUCTIONS
/ GETTING STARTED
58
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
7.1.3 PRERUN
CHECKS
This section provides a quick summary of the activities required prior to starting the
MosaiQ 125
:
1 - Ensure that all
MosaiQ 125
environmental conditions are met (refer to
chapter 18.6
).
2 - Ensure that all incubation ring insulating covers are securely in place.
3 - Ensure that all protective covers and doors are in place and securely closed.
4 - Ensure that the liquid waste canisters (red caps) are empty and the lids correctly
fitted.
NOTICE
All waste should be disposed of in accordance with your local
rules. As example: COSHH (Control of Substances Hazardous to
Health).
5 - Ensure that the connector on the liquid waste connector is securely locked in position
(refer to
chapter 7.3.10
).
6 - Ensure that the solid waste drawer is empty, firmly shut and locked into place when
replacing the biological waste bag .
(opening the solid waste drawer)
7 - Ensure that there is water at least in one canister.
8 - Prepare the MosaiQ system reagents, quality controls, MosaiQ magazines according
to their respective IFU.
7.2 GETTING
STARTED
This section explains the operations required for start-up of the
MosaiQ 125
.
7.2.1 SWITCH
ON
The following section explains how to start the
MosaiQ 125
.
1 - Ensure both
MosaiQ 125
and computer workstation are connected to the mains power
supply. For more information please refer to
chapter 5.7
of this manual.
2 - Switch on the
MosaiQ 125
(the switch is located on the right side of the
MosaiQ 125
).
(power switch localization)