QIAcube Connect MDx User Manual 12/2020
36
UV LED
The QIAcube Connect MDx is equipped with a UV LED light for decontamination. During the
maintenance decontamination procedure, the UV LED is moved over the worktable. The hood and
the waste drawer must be closed before starting the procedure and must not be opened during the
procedure.
WARNING
Risk of personal injury
[W29]
Do not expose your skin to UV-C light from the UV LED lamp.
3.4
Disposables
Rotor adapter
A disposable rotor adapter holds a QIAGEN spin column and a microcentrifuge tube in a
centrifuge bucket during sample processing. If required by the protocol, an additional column (e.g.,
QIAshredder column) can be placed in the middle position of the rotor adapter. For ease of use
and high process safety, the rotor adapters are designed so that they fit into a centrifuge bucket
only in the correct orientation. Spin-column and microcentrifuge tube lids are held securely in slots
at the edge of the rotor adapter.
The wash position of the rotor adapter is open at the bottom, enabling wash buffers to flow through
and collect at the bottom of the rotor adapter during centrifugation. The other two positions in the
rotor adapter are closed. Be sure to follow the loading instructions provided by the software.
Assembly of a rotor adapter
Microcentrifuge tube position
Middle position
Wash position
Slot for spin-column lid