The instrument does not start a load check if the black beam collimator of the ultrasonic sensor is
missing. Check if collimator has been installed before starting a load check.
Black beam collimator (see red circle) of the ultrasonic sensor
7.3.4
Touchscreen
Every time the user presses a button on the touchscreen, a small red sign is displayed at the place
where the touchscreen recognizes the contact. If the point of touch and the recognized contact are
at different positions, a re-calibration of the touchscreen can be performed. The calibration function
can be reached during the instrument’s startup procedure.
It is recommended to use a touch pen or an unused tip for optimal calibration results. In case you
use a tip, discard the tip after calibration.
To re-calibrate the touchscreen:
1.
Turn OFF the QIAcube Connect MDx.
2.
Wait a few minutes and then turn ON the instrument again.