QIAcube Connect MDx User Manual 12/2020
42
4.
To remove the foam protector around the robotic arm, gently pull the foam protector towards
you (see picture below). After removing the protector for the robotic arm, make sure to close
the QIAcube Connect MDx hood.
Foam protector for robotic arm.
4.3.2
Installation of AC power cord
1.
Remove the power cord from the foam packing material on top of the QIAcube Connect MDx.
Note
: Only use the power cord that is supplied with the QIAcube Connect MDx.
2.
Ensure that the power switch is set to OFF: outer position is OFF and inner position is ON.
3.
Check that the voltage rating on the label at the back of the QIAcube Connect MDx matches
the voltage available at the installation site.
4.
Plug the power cord into the instrument power-cord socket.
5.
Plug the power cord into a grounded power outlet.
WARNING
Damage to electronics
[W9]
Before powering ON the instrument, make sure that the correct supply voltage
is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the specifications
indicated in the type plate of the instrument.
WARNING
Electrical hazard
[W8]
Any interruption of the protective conductor (earth/ground lead) inside or
outside the instrument or disconnection of the protective conductor terminal is
likely to make the instrument dangerous.
Intentional interruption is prohibited.
Lethal voltages inside the instrument
When the instrument is connected to line power, terminals may be live and
opening covers or removing parts is likely to expose live parts.