QIAcube Connect MDx User Manual 12/2020
92
5.10
Independent centrifuge operation
The centrifuge can be operated individually if the QIAcube Connect MDx is not running a protocol.
Do not attempt to move the QIAcube Connect MDx during operation.
CAUTION
Damage to the instrument
[C7]
The QIAcube Connect MDx must not be used if the centrifuge lid is broken, or
if the lid lock is damaged.
Make sure that no loose material is inside the centrifuge during operation.
Make sure that the rotor is installed correctly and that all buckets are properly
mounted, regardless of the number of samples to be processed. Load the rotor
only as instructed by the software.
Only use rotors, buckets, and consumables designed for use with the QIAcube
Connect MDx. Damage caused by use of other consumables will void your
warranty.
We recommend replacing the centrifuge rotor and buckets after 20,000
cycles, which is equivalent to 9 years of usage with two runs per day for 220
days each year. For more information contact QIAGEN Technical Services.
WARNING
Moving parts
[W19]
In case of breakdown caused by power failure, remove the power cord and
wait 10 minutes before attempting to manually open the centrifuge lid.
CAUTION
Damage to the instrument
[C8]
After a power failure, do not move the z-module (robotic arm) manually in
front of the instrument. Damage may occur if the QIAcube Connect MDx hood
is closed and collides with the z-module.
CAUTION
Risk of overheating
[C9]
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
sides and rear of the QIAcube Connect MDx.
Slits and openings that ensure the ventilation of the QIAcube Connect MDx
must not be covered.