WARNING
Risk of electric shock
[W10]
Do not open any panels on the QIAcube Connect MDx.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
WARNING
Hazardous chemicals and infectious agents
[W16]
The waste may contain toxic material and must be disposed of properly. Refer
to your local safety regulations for proper disposal procedures.
WARNING
Risk of personal injury and material damage
[W1]
Improper use of the QIAcube Connect MDx may cause personal injuries or
damage to the instrument. The QIAcube Connect MDx must only be
operated by qualified personnel who have been appropriately trained.
Servicing of the QIAcube Connect MDx must only be performed by a
QIAGEN Field Service specialist.
WARNING
Risk of explosion
[W23]
When cleaning the QIAcube Connect MDx with alcohol-based disinfectant,
leave the QIAcube Connect MDx hood open to allow flammable vapors to
disperse.
Only clean the QIAcube Connect MDx when worktable components have
cooled down.
WARNING
Risk of fire or explosion
[W6]
When using ethanol or ethanol-based liquids on the QIAcube Connect MDx,
handle such liquids carefully and in accordance with the required safety
regulations. If liquid has been spilled, wipe it off and leave the QIAcube
Connect MDx hood open to allow flammable vapors to disperse.
WARNING
Toxic fumes
[W14]
Do not use bleach to clean or disinfect the QIAcube Connect MDx. Bleach in
contact with salts from the buffers can produce toxic fumes.
WARNING
Toxic fumes
[W15]
Do not use bleach to disinfect used labware. Bleach in contact with salts from
the buffers used can produce toxic fumes.