5.2.11
Fluorometer Input
Settings
Touch the Edit icon to view or change the settings related to the sensor.
Alarms
Low-Low, Low, High and High-High Alarms limits may be set.
Deadband
This is the Alarm Deadband. For example, if the High Alarm is 7.00, and the
deadband is 0.1, the alarm will activate at 7.01 and deactivate at 6.90.
Transmitter
Select the type of transmitter connected (2-wire loop powered, 2-wire self-powered,
3-wire, or 4-wire).
Reset Calibration Values
Enter this menu to reset the sensor calibration back to factory defaults.
Cal Required Alarm
To get an alarm message as a reminder to calibrate the sensor on a regular schedule, enter
the number of days between calibrations. Set it to 0 if no reminders are necessary.
Alarm & Datalog
Suppression
If any of the relays or digital inputs are selected, any alarms related to this input
will be suppressed if the selected relay or digital input is active. At the same time,
all datalogs and graphs containing the input will show no data for the duration of
the activation.
Smoothing Factor
Increase the smoothing factor percentage to dampen the response to changes. For
example, with a 10% smoothing factor, the next reading shown will consist of an
average of 10% of the previous value and 90% of the current value.
Max Sensor Range
Enter the value of the ppb of dye at which the sensor transmits 20 mA.
Dye/Product Ratio
Enter the value for the ratio of ppb of dye to ppm of inhibitor that is in the inhibitor
product being fed.
Name
The name used to identify the transmitter may be changed.
Type
Select the type of sensor to be connected. The choice of Analog Input is only avail-
able if that type of sensor card is installed.
5.2.12
Analog Flowmeter Input
Settings
Touch the Edit icon to view or change the settings related to the sensor.
Alarms
Low-Low, Low, High and High-High Alarms limits may be set.
Deadband
This is the Alarm Deadband. For example, if the High Alarm is 7.00, and the dead-
band is 0.1, the alarm will activate at 7.01 and deactivate at 6.90.
Totalizer Alarm
Enter the high limit on the total volume of water accumulated above which an
alarm will be activated.
Reset Flow Total
Enter this menu to reset the accumulated flow total to 0. Touch Confirm to accept,
Cancel to leave the total at the previous value and go back.
Set Flow Total
This menu is used to set the total volume stored in the controller to match the regis-
ter on the flow meter. Enter the desired value.
Scheduled Reset
Choose to automatically reset the flow total, and if so, Daily, Monthly or Annually.
Reset Calibration Values
Enter this menu to reset the sensor calibration back to factory defaults.
Cal Required Alarm
To get an alarm message as a reminder to calibrate the sensor on a regular schedule,
enter the number of days between calibrations. Set it to 0 if no reminders are necessary.
Alarm & Datalog
Suppression
If any of the relays or digital inputs are selected, any alarms related to this input
will be suppressed if the selected relay or digital input is active. At the same time,
all datalogs and graphs containing the input will show no data for the duration of
the activation.
Smoothing Factor
Increase the smoothing factor percentage to dampen the response to changes. For
example, with a 10% smoothing factor, the next reading shown will consist of an
average of 10% of the previous value and 90% of the current value.