38
Config Menu (see section 5.4)
Choose language
Touch the Configuration tab on the top of the Home screen. Touch Global Settings. Swipe up or down until the
English word “Language” is displayed and then touch it. Swipe up or down until your language is displayed and
touch it. Touch the Confirm icon to change all menus to your language.
Set date (if necessary)
In the Global Setting menu, swipe up or down until Date is displayed, and then touch it. Highlight the Day, and
then use the numeric touchpad to change the date. Touch the Confirm icon to accept the change.
Set time (if necessary)
In the Global Setting menu, swipe up or down until Time is displayed and then touch it. Highlight the digit to
change, then use the numeric touchpad to change the time. Touch the Confirm icon to accept the change.
Set global units of measure
In the Global Setting menu, swipe up or down until Global Units is displayed and then touch it. Touch the de-
sired units. Touch the Confirm icon to accept the change.
Set temperature units of measure
In the Global Setting menu, swipe up or down until Temp Units is displayed and then touch it. Touch the desired
units. Touch the Confirm icon to accept the change.
Close the Global Settings menu. Touch the Inputs tab.
Inputs (see section 5.2)
Program the settings for each input
A list of all available inputs will be displayed. Touch the S11 Input to get to the Details screen. Touch the Edit
icon. If the name of the sensor does not describe the type of sensor connected, swipe up or down until Type is
displayed. Touch the Type field. Swipe up or down until the correct type of sensor is displayed, then touch it to
highlight it. Touch the Confirm icon to accept the change. This will bring you back to the Details screen. Touch
the Edit icon and finish the rest of the S11 settings. For disinfections sensors, choose the exact sensor in the
Sensor menu. For contacting conductivity sensors, enter the cell constant. Select the units of measure. Enter the
alarm set points and alarm deadband. Set the default temperature that will be used for automatic temperature
compensation if the temperature signal becomes invalid.
When finished with S11, touch the Close icon until the list of inputs is displayed. Repeat the process for each
input.
The S12 temperature input Element should be set correctly once the S11 sensor type has been set. If not, select
the correct temperature element and set the alarm set points and alarm deadband. Generic, ORP and disinfection
sensors do not have temperature signals and are preset to Unassigned.
To calibrate the temperature, return to the S12 Details screen, touch the Calibrate icon, and touch the Enter icon
to perform a calibration. If either input card is a Dual Analog Input card (4-20mA signal), then select the type of
sensor that will be connected. Select Fluorometer if a Little Dipper 2 will be connected. Select AI Monitor if the
device can be calibrated on its own and the
EP2100
™ calibration will only be in units of mA. Select Transmit-ter
if the device connected cannot be calibrated on its own and the
EP2100
™ will need to be used to calibrate in
engineering units of measure.
If a flow switch or liquid level switch is connected, D1 through D12 (whichever one has the device connected
to it) should be set to DI State type (if no switch is connected, select No Sensor). Set the state that will possibly
interlock control outputs (refer to the Outputs settings to program which outputs, if any, will be interlocked by
the switch). Set the state, if any, that will result in an alarm.
If a contacting head or paddlewheel flow meter is connected, D1 through D12 (whichever one has the device
connected to it) should be set to that type (if no flow meter is connected, select No Sensor). Set the units of mea
-
sure, volume/contact or K factor, etc.