14
Error describtion
Measures
1.
The device indicates a permanent
overflow.
• The input has a very large measured value,
check the measuring distance.
• If an input with a small sensor signal is selected,
it is only connected on one side or the input is
open.
• Not all activated interpolation points are
parameterized. Check whether the relevant
parameters are set correctly.
2.
The device indicates a permanent
underflow.
• The input has a very small measured value,
check the measuring distance.
• If an input with a small sensor signal is selected,
it is only connected on one side or the input is
open.
• Not all activated interpolation points are
parameterized. Check whether the relevant
parameters are set correctly.
3.
The device displays
„
HELP
“ in the
7-segment display
• The device has detected an error in the
configuration memory, perform a reset to the
default values and reconfigure the device
according to your application.
4.
Programming numbers for
parameterization of the input are not
available
• The programming lock is activated
• Enter correct code
5.
The device displays „
Err1
“ in the
7-segment display
• In case of errors of this category please contact
the manufacturer.
6.
The device does not respond as
expected.
• If you are not sure that the device has been
parameterized before, restore the delivery state
as described in
chapter 6
.
10. Error description
10. Error description