5. Setting up the device
Menu level
Parameterization level
Setting the tara/offset value,
TArA:
Default:
0
The specified value is added to the linearized value. This is how the characteristic curve can be
shifted by the selected amount.
Zero point slowdown,
ZErO
:
Default:
0
With zero point calming, a value range around the zero point can be preselected at which the
display shows a zero. If, for example, a 10 should be set, the display would show a zero in a
value range from -10 to +10 and continue with -11 below and +11 above.
Assignment (deposit) of key functions,
Tast
:
Default:
no
Here, either a min/max value query or a limit value correction can be stored on the direction
keys for the operating mode.
If the min/max memory is activated with
EHER
the measured min/max values are stored during
operation and can be retrieved via the direction keys [
▲] [▼]. The values are lost when the
device is restarted. The values are lost when the device is restarted.
If the limit value correction
LI.1
is selected, the values of the limit values can be changed during
operation without interfering with the operating sequence. With
tara
the display is tared to zero
and permanently stored as offset. The display acknowledges the correct taring with
0000
in the
display. If
No
is parameterized, the direction keys [
▼] [▲] have no function in the operating
mode.
Flashing on display,
FLAS:
Default:
no
Here a display flashing can be added as an additional alarm function either to the first limit
value (selection:
LI-1
), to the second limit value (selection:
LI-2
) or to both limit values (selection:
LI-12
). With
No
(factory setting) no flashing is assigned.
Limit value,
LI-1:
Default:
0200
The limit value specifies the threshold above which the alarm reacts or is activated /
deactivated.