Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
35
www.promega.com
TM320 · Revised 4/15
6.
Remove the black LEV Plunger Bar Adaptor
by
pulling down on the LEV Plunger Bar Adaptor to
re-
lease the magnets that hold it in place.
Note:
Pull down at an angle so that the back
of
the magnet is released first.
7.
Insert the SEV Magnetic Rod Assembly. The set screws
should be facing toward you. Press firmly to place the
Magnetic Rod Assembly in position. It will fit tightly
into place.
8.
Secure the SEV Magnetic Rod Assembly with the three thumbscrews, and hand-tighten.
9.
Close the door; the screen will indicate that the hardware has been changed. The instrument indicates a
hardware change once the door is closed, whether or not you actually completed the physical hardware
change. Confirm that the hardware mode shown matches the physical hardware installed. If the screen does
not match the physical hardware installed, turn the instrument o
ff
, install the appropriate hardware and
restart the machine. The hardware mode indicated on-screen should match the physical hardware installed.
Important.
When the
Home
screen is accessed, it will indicate the new hardware mode. Always confirm
that the hardware mode shown matches the physical hardware installed. Running the instrument when the
hardware mode shown does not match the physical hardware installed will damage the instrument.
6026TB
6024TB
!