Review the scanned data. All three codes should match. If they do not match, you can reselect the option to
change and rescan the information.
5.
Scan the kit box bar code label. The bar code data will appear on the screen next to the
Kit Info
box. If the
kit expiration date has passed, an error message will be displayed, and a di
ff
erent kit lot must be used. If
more than one kit lot is to be used, the new kit lot number can be scanned when the first cartridge and
elution tube from the new kit are added to the cartridge rack.
Note:
The kit lot number will automatically copy into the next sample position. If a new kit lot number is
used, delete the previous lot information and scan the new information or enter it manually.
6.
On-screen data can be removed from each option by highlighting the option and selecting the
Clear Selection
option. All of the data can be removed by selecting the
Clear All
option.
7.
Select
Next Sample
to add the information for the next sample. For empty sample positions, select
Next Sample
repeatedly until the required sample position is reached.
8.
Repeat Steps 2 to 5 for the new sample and associated cartridge and elution tube, and place them in the
required positions. Select the
Next Sample
option. Repeat until all samples are loaded.
9.
If there are fewer than 16 samples, select
Final Sample
when all of the samples are loaded. This will open
the
Verify Scanned Data
screen. This screen allows you to review the bar code data entered.
8262T
A
10. If any incorrect data are identified for a sample, you can use the navigation buttons to return to the
Barcode Input
screen where Position 01 will be shown. Select
Next Sample
to navigate to the required
sample, and re-enter all of the bar code information.
11. For SEV methods, once all of the data have been verified, press the Run/Stop button and close the door to
start a run.
12. For LEV methods, once all of the data have been verified, place the cartridge rack into the instrument, close
the door, and press the Run/Stop button.
Figure 24. The Verify Scanned Data screen.