7.A. Changing the Hardware (continued)
6.
Insert the blue LEV Magnetic Rod Assembly in the correct
orientation. The screws should be facing toward
you. Press firmly to place the LEV Magnetic Rod Assembly
in position. It will fit tightly in place.
7.
Secure the LEV Magnetic Rod Assembly with the three
thumbscrews from the SEV Magnetic Rod Assembly, and
hand-tighten.
8.
Close the door; the screen will indicate that the hardware
has been changed. The instrument indicates a hardware
change once the door is closed whether or not you actually
completed the physical hardware change. Confirm that the
hardware mode shown matches the physical hardware installed. If the screen does not match the physical
hardware installed, turn the instrument o
ff
, install the appropriate hardware and restart the machine. The
hardware mode indicated on-screen should match the physical hardware installed.
Important
. When the Home screen is accessed, it will indicate the new hardware mode. Always confirm
that the hardware mode shown matches the physical hardware installed. Running the instrument if the
hardware mode shown does not match the physical hardware installed will damage the instrument.
Changing from LEV to SEV Hardware
1.
Select
Hardware
on the
Maxwell
®
16 Configuration
screen.
2.
Follow the on-screen instructions. First, open the door.
3.
The Magnet Assembly will move down to allow you to remove the Magnetic Rod Assemby. Do Not Push on
the Magnetic Rod Assembly.
4.
Unscrew the three thumbscrews on top of the
magnetic rod assembly.
5.
Gently lift the LEV Magnetic Rod Assembly, and
remove it.
Orientation of the Magnetic Rod Assembly.
!
9514TB
Back
Front