Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
49
www.promega.com
TM320 · Revised 4/15
5.
Lysis Time
. Select the down arrow, and add the new value using the keypad. The maximum time that can be
added is 60 minutes; the minimum is zero minutes.
Binding Time.
Select the down arrow, and add the new value using the keypad. The maximum time that can
be added is 60 minutes; the minimum is zero minutes.
Drying Time.
Select the down arrow, and add the new value using the keypad. The maximum time that can
be added is 60 minutes; the minimum is zero minutes.
Elution Time.
Select the down arrow, and add the new value using the keypad. The maximum time that can
be added is 60 minutes; the minimum is zero minutes.
Number of Cycles.
Select the down arrow, and add the new value using the keypad. The maximum number
of cycles that can be added is 3; the minimum is 1.
6.
Once the user method parameters are defined, the new method can be named and a PIN added. PIN protec-
tion can prevent inadvertent changes. After changing the parameters, press the Run/Stop button to open the
User Protocol Options
screen.
8237T
A
7.
To name a method, select
Edit/Add Protocol Name,
and use the keyboard to enter the method name. Select
the Run/Stop button to return to the
User Protocol Options
screen.
8.
A PIN, which must be entered to edit the method, can be added. The PIN can contain up to four digits.
We recommend that you keep a separate list of all methods and associated PINs. In the
U
ser Protocol
Options
screen, select the
Edit/Add PIN
option. Using the keypad, enter the PIN, and then press the
Run/Stop button. Confirm the PIN, and press the Run/Stop button again. An administrative password is
included in the Welcome Letter, received with the instrument, which is used to access all PIN-protected op-
tions.
The Maxwell
®
16 MDx Instrument allows storage of 10 custom SEV and 10 custom LEV methods. If there
are 10 existing custom methods, more custom methods can only be added by changing an existing method.