5.C. Performing a Sample Purification Run
Follow the on-screen instructions for SEV or LEV setup.
8219T
A
8220T
A
Once all of the samples, plungers and elution tubes have been loaded, press the Run/Stop button and close the
door. The run will begin automatically.
Important.
The plungers must be placed in the correct starting position. If the instrument goes through a run
without plungers, the magnetic rod assembly must be removed and cleaned (see Section 8.B). The cartridges
containing the samples must be discarded.
During the run, the
Processing
screen will display the current run step and run time remaining. The instrument
will automatically perform periodic self-diagnostic tests to verify that the platform, plunger bar and magnetic rod
assembly are within calibration during the purification procedure. If the instrument detects that any of these parts
are not within calibration during a run, the method will pause and display
Calibration Error
.
8221T
A
Figure 25. The SEV Set Up screen.
Figure 26. The LEV Set Up screen.
Figure 27. The Processing screen.
!