10.D. Service (continued)
In the
Utilities Menu
screen, select
Service Report
to view and print a current service report. The report will list
the report date, the Instrument Serial Number, the date last service and preventative maintenance were per-
formed, the number of run cycles the instrument has performed, the total number of hours the UV lamp has run
and the firmware version.
You also can print the service report using the Maxwell
®
Sample Track Software. See Technical Manual #TM314,
which describes how to do this.
The
Utilities Menu
screen also allows the following options:
•
Run Count
shows the number of times the instrument has been run.
•
UV Bulb Time
shows the number of hours the bulb has been run. If you are replacing the bulb (see Section
8.E.), you may reset the number of hours to zero by selecting the Yes option in the UV bulb screen.
•
Reset Name
allows you to rename the instrument. Use the keyboard to add the new name.
•
Reset Wizard
opens the
Reset Wizard
screen.
8228T
A
If you select
Wizard Only
, after confirmation the Wizard will reset, and when the instrument restarts, the Wizard
begins.
Selecting
All Settings
will remove all custom settings as well as reset the Wizard.
These are the parameters that can be changed in your facility. For any other service, see the Warranty information
in Section 11.
Figure 45. Reset Wizard screen.