4.E. Operational Setup (continued)
2.
Touch Screen.
You will need to calibrate the TouchScreen after shipping.
8205T
A
8206T
A
The “
Touch Screen Reset Successful
” message indicates that the calibration was successful, and the Wizard can
continue.
Figure 12. Calibrating the Touch Screen.
To calibrate the screen
touch the “+” symbols in the order they are presented. A confirmation
screen will open.
Figure 13. Confirmation screen for the Touch screen.
Touch all
four boxes; if they all change color, press the Stop/Run button on the
instrument keypad. If they do not all change color, press the back arrow
button and repeat the calibration and confirmation.