4.E. Operational Setup (continued)
The Wizard will indicate that it is finished and will restart the instrument. Upon restart the instrument will
perform a self-diagnostic test before opening the
Home
screen. The platform, plunger bar and magnetic
rod assembly are moved to check operation, and the UV light is briefly switched on to ensure the bulb is
functioning.
During the diagnostic test the firmware version is displayed on the screen. If you need to update to a new
firmware version, see Section 10.
If you have chosen to run the UV lamp each time the instrument is switched on, the UV lamp will come
on and run for the specified time following a verification screen. You will have the option to cancel this
treatment.
Important
. Do not open the door if the UV lamp is on. The UV lamp will not run if the door is open.
The
Home
screen contains the menu options that allow access to all of the functions available for the
instrument and displays the hardware mode (SEV or LEV).
8286T
A
8213T
A
Figure 19. Home screen.
Selecting
Run
allows a
purification run to be set up and performed (see
Section 5). Selecting
Data Transfer
allows the
transfer of run data from the instrument to a printer
or computer if you are using the Maxwell
®
Sample
Track Software (see Section 6), or it allows deletion
of run data from the instrument. Selecting
Setup
opens the
Maxwell
®
16 Configuration
screen.
Figure 20. Maxwell
®
16 Configuration screen.
The functions to change the hardware (
Hardware
) and
firmware (
Configuration
) settings (see Section 7) are
accessible from this screen. Also, the UV lamp may be
run from this screen
(
UV On
; Section 8). This screen
also provides access to the Service functions (
Service
Mode
; Section 10.D). User protocols also may be set up
from this screen by selecting
User Protocol
(Section
10.A).
!