35
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
Methods
The ‘Methods’ screen displays a list of the currently installed methods in the user interface
including the name, version number and catalog number for each method. From the ‘Methods’
screen the administrator can view, import and delete methods from the Maxwell
®
CSC 48
software (Figure 25).
1. If you desire to view the methods currently installed in the Maxwell
®
CSC 48 software, touch
the
Methods
button from the ‘Administrator Page’ screen.
2. Touch the
Back
icon in the upper left corner of the screen to return to the ‘Administrator
Page’ screen.
Importing and Deleting Methods
As Promega provides new purification kits for use with the Maxwell
®
CSC 48 instrument, new
extraction methods can be added to the Maxwell
®
CSC 48 software. Occasionally, an existing
extraction method may have to be updated. Only administrators can add new extraction
methods or delete or update existing methods. Administrators can download new or updated
extraction methods from the Promega web site at:
www.promega.com/resources/software-firmware/maxwell-maxprep/
maxwell-csc-48-methods/
Methods are provided as files with a .package extension. Follow the instructions below to import
a method into the Maxwell
®
CSC 48 software.
1. Save the method (.package) file on the Tablet PC attached to the Maxwell
®
CSC 48.
2. If not already running, start the Maxwell
®
CSC 48 software (in either IVD Mode or RUO
Mode) by double-touching the software icon on the desktop.
3. Navigate to the ‘Methods’ screen (‘Home’
→
Settings
→
Administrator
→
Methods
).
4. Touch the
Select Import Package
button on the bottom right side of the screen to open
the ‘Select File’ browser screen.
5. Within the ‘Select File’ screen, navigate to the location at which you saved the method
(.package) file. The current path will be shown in yellow boxes along the top of the window.
Touch any aspect of the path to navigate to the desired folder. Touch the
Drive\
button to
select the drive in which to search. Folders within the current path are shown as red
rectangles on the left side of the screen. Valid method (.package) files will be shown as blue
rectangles (yellow when selected) on the right side of the screen.
6. Click the blue rectangle(s) corresponding to the desired (.package) file(s) to highlight it, then
press the
OK
button.
Maxwell® CSC 48 User Interface