28
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
5.2 User Interface Settings (continued)
Export Logs
The
Export Logs
button will export instrument logs for troubleshooting purposes. If in the
course of troubleshooting an instrument problem you are directed to export logs to send to
Promega Technical Services, touch this button to generate instrument-specific log files.
1. Touch the
Export Logs
button to display the ‘Export Folder’ screen (Figure 19, Panel A). This
screen allows the operator to select the path to which the instrument logs should be exported.
(The default export path for instrument logs is C:\Exports\.)
2. Using the Yellow and Red rectangle buttons you can select the folder location where the
instrument logs will be saved. The current path is indicated by the yellow rectangles at the
top of the ‘Export Folder’ screen. Any folders present within the selected directory are
displayed as red rectangles in the main portion of the screen. Touch the
Drive\
button to
navigate to the drive location of the desired folder. Touch red folder buttons to navigate to
the desired folder location.
3. Once a path has been defined, touch the
Save
button to export logs to the specified path,
or touch
Cancel
to return to the ‘Settings’ screen without exporting logs.
4. After pressing
Save
, the ‘Export’ screen will be displayed, indicating that the log files were
successfully exported to the path that was specified (Figure 19, Panel B).
5. On the ‘Export’ screen you can press
Open
to open a Windows File Browser showing the
folder location of the exported files. Press
Done
to close the ‘Export’ screen and return to
the ‘Settings’ screen.
Instrument logs will be exported as a zip file to the path that was specified by the operator. The
zip file of the logs should be retrieved from the specified location and sent to a Promega
Technical Services representative for further troubleshooting.
A.
B.
Figure 19. Export Logs. Panel A.
The ‘Export Folder’ screen is displayed after touching the
Export Logs
button. Specify a path to which the instrument log files should be exported. Touch
Save
to export instrument logs to the specified location or
Cancel
to return to the ‘Settings’
screen without exporting logs.
Panel B.
After instrument log files are exported, the ‘Export’
screen is displayed, indicating that logs have been saved to the specified path. Touch
Open
to view the folder location of the exported instrument log files. Press
Done
to exit the ‘Export’
screen and return to the ‘Settings’ screen.
14413T
A
14414T
A
Maxwell® CSC 48 User Interface