Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
5.3 Administrator Settings (continued)
2. Three sanitization settings are present on the ‘Sanitization Settings’ screen (Figure 24).
These are described below:
a. Default sanitization duration: This setting defines the duration of the UV treatment (in
minutes) performed when the
Sanitization
button on the ‘Home’ screen is pressed.
b. Sanitize after extraction for (
optional
): This setting defines the duration of UV treatment (in
minutes) that will be automatically performed after completion of an extraction method run.
c. Sanitize on software start-up for (
optional
): This setting defines the duration of UV
treatment (in minutes) that will be automatically performed when the Maxwell
®
CSC 48
software is started.
3. Touch the text box next to "Default sanitization duration:" to open the on-screen number
pad. Enter the number of minutes to be used for UV sanitization when the
Sanitization
button on the ‘Home’ screen is pressed. Touch the
OK
button on the on-screen number
pad to accept the duration value or touch the
Cancel
button on the on-screen number pad
to discard changes.
4. To enable either of the additional UV sanitization options, touch the checkbox next to the
desired option. A checkmark will be visible in the checkbox next to the enabled option, and
the text box associated with the option will become active. Touch the text box associated
with the desired option to open the on-screen number pad. Enter the number of minutes of
UV sanitization to be performed for the desired option. Touch the
OK
button on the on-
screen number pad to accept the duration value, or touch the
Cancel
button on the
on-screen number pad to discard changes.
5. Once all UV sanitization options have been set, touch the
Save
button to accept and save
the settings. To discard any changes to the UV sanitization options, touch the
Cancel
button. Touching either button will result in a return to the ‘Administrator Page’ screen.
Figure 24. The ‘Sanitization Settings’ screen.
14417T
A
Maxwell® CSC 48 User Interface