Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
6.2 Starting a Method (continued)
User Aborts Method
If the method is aborted by the user, the ‘Protocol Running’ screen will indicate that the
method has been aborted (Figure 47). After a method is aborted, the Current Step will be
listed as Aborted by user. After aborting the method, press the
Open Door
button. The Vision
system (if enabled) will determine whether plungers have been unloaded successfully, and if not,
will attempt to unload them. Otherwise the ‘Clean Up’ screen (Figure 48) will be displayed.
Figure 47. ‘Protocol Running’ screen after abort by the user.
If a method is aborted by
the user, the ‘Protocol Running’ screen will display a progress bar of 100%, and Current Step is
shown as Aborted by user. After a method is aborted, press the
Open Door
button.
Instrument Error
If the method is aborted due to instrument error, the ‘Protocol Running’ screen will change to
indicate that the method has been aborted and will display an error message.
After a method is aborted, the Current Step will list the reason for aborting the method.
An aborted run
(initiated by the user
or due to instrument
error) will result in
the loss of all the
samples. Do not
attempt to repurify
samples from an
aborted run.
16984T
A
An aborted run
(initiated by the user
or due to instrument
error) will result in
the loss of all the
samples. Do not
attempt to repurify
samples from an
aborted run.
Operating the Maxwell® CSC 48 Instrument